| About ROSAT |
ROSAT Home Page | ROSAT Images |
|---|
For every requested pointing direction a target form must be filled out.
).
) will be generated from the target form.
The ``Additional Constraints'' entry indicates whether special scheduling
constraints
(i.e., constraints for time-critical observations) apply or not. If
this entry is ``Y(es)'', an ``Observational Constraint Form'' must
accompany the source form. Note that the number of observations can only
be larger than one if additional scheduling constraints exist.
The ``WFC Zoom On'' entry requests a small or large WFC field of view
(cf., § 5.3
for details); it is strongly suggested to
use the ``NOZOOM'' option for all filters unless there are very specific
reasons to do otherwise.
The requested instrument configuration
is determined by assigning an instrument code to each of the three
available instruments (HRI, PSPC and WFC) in the following way:
The primary instrument is denoted by the entry ``1'', the secondary instrument
by the entry ``2'', and the instrument not to be used by the entry ``0''.
The various possible combinations are listed in Table 9.1
.
Note that Principal Investigators from the US must not select the WFC as
primary instrument and that PSPC and HRI are mutually exclusive.
| PSPC | HRI | WFC | Comments | |
| PSPC+WFC | 1 | 0 | 2 | |
| HRI+WFC | 0 | 1 | 2 | |
| WFC+PSPC | 2 | 0 | 1 | only non-US proposers |
| WFC+HRI | 0 | 2 | 1 | only non-US proposers |
For both the PSPC and the WFC, different filter settings can be chosen
(see Chapter 4
and Chapter 5
). However the number
of filter changes during the
execution of any ROR is limited to one for the PSPC. WFC filter changes
more frequent than about once per 1000 s should not normally be
requested; for a limited number of observations a higher frequency
may be permitted if judged to be of sufficient scientific merit.
The requested filter sequences can be specified in items 4. PSPC
Filter Sequence and 5. WFC Filter Sequence below
and will be scheduled - if at all possible
- in the order as specified by the guest investigator with the
percentages (in terms of total requested time) specified in the target form.
It is important to note that the specified sequence implies a priority
order. If the total time allocated to the proposal is reduced for
any reason (for example, by the decision of the selection committee), the
observation will be carried out with the times allocated to each filter
according to the requested percentages, unless the time implied falls
below the minimum time requested. In this case the specific filter
will not be selected and the total time will be distributed among the
remaining filters according to the percentages specified. Observers
who wish not to run the risk of an observation with any
particular filter being dropped out should specify a minimum useful
time of zero for all filters in the minimum time entry.
US proposers (who have no proprietary
data rights to the WFC) are encouraged but not requested to fill out
the WFC items.
for details) to indicate the filters
of your choice. If only a single filter
is desired provide the appropriate filter identification with
percentage time set to 100 percent and minimum time set to zero.