ME CALIBRATION AND UPDATES TO THE CCF
Introduction
Changes in both format and content of the ME CCF have been introduced
and are incorporated on FOTs produced from about the end of April
1985. Users can identify the new CCF by layout numbers of 5 and 3 for
data types CB and SO respectively. The only other data type to have
been changed is type SU.
These changes reflect recent work on both the Ar and Xe detector
calibrations. Fitting the combined Ar and Xe half 1 spectra for the
1984 day 318 Crab Nebula observation using the Observatory Interactive
Analysis System gives a chisq. of 141 for 148 degrees of freedom
(Figure 1), with the column density fixed at 3x1021
atoms/cm2 and the
power law slope at -2.1. 1% systematic uncertainties were included for
both detectors. A normalisation figure of 9.53 and relative
normalisation between the Ar and Xe detectors (a free parameter in the
fitting) of 0.998 were determined. The improved Ar calibration results
in residuals a factor of 2-4 smaller than in the previous calibration
although the characteristic shape, a dip between 2-3 keV, is still
present. For some unexplained reason detector E appears to give
systematically larger residuals than the other detectors.
Changes to the Ar calibrations concern mainly the gain curves and
their time dependence. The calibrations to date have assumed that the
observed gain changes could be modeled as a change in the overall
gain of each detector. Analysis of 6 Crab Nebula observations suggests
that this is an oversimplification and that the amount of curvature,
or non-linearity, is also time-dependent. To include this effect, the
gain change parameters for both A1 and A2 are now included on the
CCF.
The main change to the Xenon calibration is that an extra term has to
be included in the Xenon resolution function, broadening the
resolution by about 10% at 20 keV. This additional broadening, which
was not present in the ground calibrations, is possibly caused by
noise on the system power supplies. Its effect is expected to be
considerably smaller for Ar. In addition there is evidence for a
gradual decrease in overall gain of the Xe detectors, corresponding to
a gain change of about 3.7% since launch. The change appears to be
linear with time and occurs over the whole mission, except for
detector A which exhibits an apparent change in gradient around
1984(85). Note that the Xe gain changes are in the opposite sense to
Argon detector gain increases (ref. Express No.9, p.38).
Revised Record Layouts
Data Type CB
Words | Contents |
1-128 | As before |
129-144 | Ar Fiducial or Reference Times dets A to H. |
145-160 | Start Time 1 of Applicability for A1 change |
161-176 | Start Time 1 of Applicability for A2 change |
177-184 | A1 Gain change/day (*E-4) at Times T1 to T2. |
185-192 | A2 Gain change/day (*E-5) at Times T1 to T2. |
193-200 | Amplifier Gain Settings at Fudicial Time |
201-216 | Start Time 2 of Applicability for A1 change |
217-232 | Start Time 2 of Applicability for A2 change |
233-240 | A1 Gain Change/day (*E-4) at Times T2 |
241-248 | A2 Gain Change/day(*E-5) at Times T2 |
249-384 | Currently Blank |
385-504 | As words 129-248 but for Xe except that the
units of the A1 and A2 change are E-7/day |
505-768 | Currently Blank |
Note that if Start Time 2 = 0 then the Gain Change parameters for Time 1 apply at all times.
Data Type SD
Words | Contents |
1-128 | Unchanged layout, Ar data |
129-129 | N1/0.01 for Xe Det. A Mode A |
130-130 | N2/0.01 for Xe Det. A Mode A |
131-131 | N3/1.E-4 for Xe Det. A Mode A |
132-132 | N4/1.E-5 for Xe Det. A Mode A |
133-133 | N5/0.01 for Xe Det. A Mode A |
134-134 | N6/0.01 for Xe Det. A Mode A |
135-135 | N7/1.E-4 for Xe Det. A Mode |
136-136 | N8/1.E-2 for Xe Det. A Mode |
137-144 | As above but for Mode B (The nominal) |
145-160 | As words 124-144 but for Det. B |
161-176 | As words 124-144 but for Det. C |
177-192 | As words 124-144 but for Det. D |
193-208 | As words 124-144 but for Det. E |
209-224 | As words 124-144 but for Det. F |
225-240 | As words 124-144 but for Det. G |
241-256 | As words 124-144 but for Det. H |
How to use the Revised Calibrations
1. Xenon Resolution Function
The spectral distribution function for Ar is unchanged. For Xe it is now given by:
Del.V/V= SQRT (R2+(N8x2O/E)2)
where E is energy in keV and R the original Resolution. See the FOTH ch. 3.7.2.6, ie:
R=N1/10+N2/100/E+N3/l.E4xE+N4/l.E5xE2, E < N5/100
R=N6/100+N7/l.E4xE, E > N5/100
The values of N1 through N8 are given in the CCF record type SD.
2. Xenon Gain Coefficients
The Xe gain coefficients at time can be determined using the
following relations where T2 is the start of applicability of the
second set of gain change parameters, Tfud is the Fiducial time,
DEL(T) the gain change per/day at time interval T, and A2 the gain
coefficient.
DX = 0.0
IF (t.GT.T2.AND.T2.NE.0) then
DX = (t-T2)*DEL(T2)
t=T2
ENDIF
TX = t-T1
DG = TX*DEL(T1)+DX
A2 = A2/(DG*1.0)
A2 must be corrected for any difference in digital amplifier
settings MAIN) between the observation time and the fiducial time
(see